Warning:
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves while installing reservoir bottles.
To install the reservoir bottles:
1.
Remove the reservoir bottle caps.
2.
Screw the reservoir bottles onto the instrument, as shown below.
3.
For each reservoir bottle, ensure that the end of the solvent delivery
tube is positioned so that it is close to, but does not touch, the bottom of
the bottle.
To install low-volume vials:
Warning:
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves while installing low-volume vials.
1.
If a standard reservoir bottle is fitted, remove it.
2.
Screw the low-volume adaptor into the manifold and tighten it.
3.
Screw the low-volume vial into the adaptor.
4.
For each low-volume vial, ensure that the end of the solvent delivery
tube is positioned so that it is close to, but does not touch, the bottom of
the vial.
To avoid personal contamination with biohazards or
To avoid personal contamination with biohazards or
March 14, 2014, 715004387 Rev. B
Preparing the IntelliStart Fluidics system
TP03410
53