Waters Xevo TQD Overview And Maintenance Manual page 225

Hide thumbs Also See for Xevo TQD:
Table of Contents

Advertisement

NW40 clamps (included in the installation kit)
NW25/NW40 adaptor (included in the installation kit)
10-mm reducer fitting (included in the installation kit)
10-mm ID nylon tube (included in installation kit)
12.7-mm clear PVC exhaust tubing (included in the Waters Rotary
Pump Kit)
PVC hose clamps (included in the Waters Rotary Pump Kit)
1-inch ID vacuum hose (included in the Waters Rotary Pump Kit)
Bear in mind these requirements when connecting the roughing pump:
The pump must be horizontal, or within one degree of horizontal.
The ambient temperature of area in which you place the pump must
range between 15 and 40 ºC.
The pump must be oriented so that it permits easy access to the gas
ballast valve and oil-level sight glass.
For proper ventilation, the following minimum clearances must apply:
Back-side minimum clearance
is 15.24 cm (6 inches)
Left-side minimum clearance
is 15.24 cm (6 inches)
Connecting the Edwards oil-free roughing pump
Right-side minimum clearance
is 15.24 cm (6 inches)
TP03165
Front-side minimum clearance
is 35.56 cm (14 inches)
March 14, 2014, 715004387 Rev. B
225