4 Maintenance Procedures
To fit the sampling cone assembly to the source:
1.
Ensure that the source isolation valve is in the closed position (see
page
96).
2.
Hold the sampling cone assembly so that the cone gas nozzle handle is
oriented horizontally and at the top, and then slide the sampling cone
assembly into the ion block assembly.
Ion block assembly
3.
Grasp the cone gas nozzle handle and use it to rotate the sampling cone
assembly 90 degrees, moving the handle downward from the horizontal
to the vertical position.
4.
Open the source isolation valve (see
5.
Close the source enclosure.
118
March 14, 2014, 715004387 Rev. B
TP03132
Sampling cone assembly
page
97).