Installing the combined APPI/APCI source
Required material
Chemical-resistant, powder-free gloves
Warning:
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves while performing this procedure. The source
components can be contaminated.
Warning:
prepare the instrument according to the procedure on
"Preparing the instrument for operations on or inside its
To install the combined APPI/APCI source:
1.
Prepare the instrument for working on the source (see
Warning:
with the probe and source; these components can be hot.
2.
Remove the probe from the currently installed source.
•
If you are removing an ESI probe, see
•
If you are removing an IonSABRE II probe, see
3.
Remove the existing source enclosure (see
4.
Install the combined APPI/APCI source enclosure (see
5.
Install the corona pin (see
6.
Connect the APPI drive cable to the instrument's front panel connector.