15. Carefully push the PEEK union/UNF coupling assembly and capillary
into the probe assembly so that the locating pin on the UNF coupling is
fully engaged in the locating slot at the head of the probe assembly.
UNF coupling locating pin
16. Fit the nebulizer adjuster knob to the PEEK union/UNF coupling
assembly.
17. Finger-tighten the nebulizer adjuster knob onto the probe assembly.
18. Fit the probe gasket and end-cover to the probe assembly.
19. Use the combined 2.5-mm Allen wrench and cone extraction tool to fit
and tighten the 3 screws retaining the probe end-cover.
20. Return the combined 2.5-mm Allen wrench and cone extraction tool in
its storage location on the source adaptor housing.
Notice:
• To avoid damaging the probe heater, take great care to grip the
heater so as not to damage its electrical wiring.
• To avoid damaging the probe heater's electrical connections,
capillary sleeve, or capillary, take great care when fitting the
heater over the capillary sleeve.
21. Fit the probe heater (see step 1 through step 3 on
22. Fit the probe to the instrument (see
23. In the Instrument Console, click API
desolvation gas flows.
Replacing the IonSABRE II probe sample capillary
Probe assembly locating slot
page
63).
to start the probe and
March 14, 2014, 715004387 Rev. B
page
169).
165