4 Maintenance Procedures
4.
Use the needle-nose pliers to slide a new ferrule onto the capillary.
5.
Insert the capillary in the PEEK union, and ensure that it is fully
seated.
6.
Screw the UNF coupling into the PEEK union, finger-tight only.
7.
Gently tug on the capillary, testing to ensure that it stays in place.
8.
Use the 7-mm wrench to tighten the locknut against the PEEK union.
Warning:
wear safety goggles when performing the leak test.
9.
Perform a leak test by attaching the free end of the PEEK tubing to an
LC pump and pumping 50:50 acetonitrile/water through it at 1 mL/min.
If leakage occurs, disassemble and remake the connection, and
repeat the leak test.
If the backpressure on the LC pump is high, replace the capillary,
and repeat the leak test.
10. When no leakage occurs and the backpressure on the LC pump is
normal, disconnect the PEEK tubing from the LC pump.
11. Remove the probe inlet connector and PEEK tubing from the PEEK
union.
12. Remove the probe heater (see step 2 on
13. Fit the PEEK union/UNF coupling assembly to the nebulizer adjuster
knob.
14. Carefully thread the capillary through the probe assembly.
164
March 14, 2014, 715004387 Rev. B
To avoid eye injury from high-pressure liquid jet spray,
page
167).