Waters Xevo TQD Overview And Maintenance Manual page 49

Hide thumbs Also See for Xevo TQD:
Table of Contents

Advertisement

UNIFI software:
a.
From the Xevo TQD console, in the Maintain pane, click Vacuum.
b.
On the Status page, click Pump.
After a 20-second delay, during which the turbopump is starting,
Tip:
the roughing pump starts. The Instrument System pane shows that
the instrument is in Standby mode, and the Operate LED remains off.
c.
Wait a minimum of 2 hours for the instrument to be fully pumped
down (evacuated).
In the Instrument Summary pane, the Status indicator shows
Tip:
green when the instrument is fully pumped down (evacuated).
d.
Click Instrument Operate Mode
When the mass spectrometer is in good operating condition,
Result:
UNIFI software displays status "Running" in the Instrument
Summary pane.
If clicking Instrument Operate Mode fails to put the instrument
Tip:
into Operate mode, UNIFI software displays corrective actions.
Starting the mass spectrometer
.
March 14, 2014, 715004387 Rev. B
49