6.
Holding the oil demister element slightly elevated to prevent the loss of
the spring, remove the exhaust flange from the oil demister element.
7.
Remove the spring from the oil demister element.
Warning:
oil demister element in accordance with local environmental
regulations. It can be contaminated with irritant oil, or with
biohazardous or toxic materials.
8.
Dispose of the oil demister element in accordance with local
environmental regulations.
Replacing the roughing pump's oil demister element
To avoid spreading contamination, dispose of the