4.
Open the air filter grill by pulling the tab at the top of the grill toward
you.
Air filter grill tab
5.
Remove and dispose of the old filter.
6.
Place the new filter flat on the inside part of the grill, with its edges
beneath the metal lip.
7.
Close the air filter grill.
8.
Connect the probe cable to the high-voltage connector.
9.
Close the source enclosure.
10. Lower the instrument's visor.