3 Changing the Mode of Operation
5.
Unscrew the union from the end of the sprayer assembly.
6.
Remove the existing capillary from the sprayer.
7.
Carefully remove the new borosilicate glass capillary from its case by
lifting vertically while pressing on the foam with two fingers.
Foam
8.
Load sample into the capillary using a fused silica syringe needle or a
GELoader tip, minimizing any bubbles between the capillary tip and the
sample.
78
March 14, 2014, 715004387 Rev. B
Capillary
Union
Capillary