Connecting the workstation

Before connecting the workstation to the instrument, set up the workstation
according to its accompanying instructions. Locate the workstation within
5 meters (16 feet) of the instrument.
Requirement:
susceptibility to radio and electrical frequency interference.
To connect the workstation:
1.
Connect the monitor to the PC.
2.
Connect one end of the shielded network cable to the appropriate port on
the rear panel of the instrument.
3.
Connect the other end of the shielded network cable to the port labeled
Instrument LAN on the workstation rear panel.
To connect the instrument to the power source:
Requirement:
you complete the installation procedures in the previous sections.
1.
Select the correct power cord for your location.
2.
Connect the female end of the power cord to the power port on the rear
panel of the instrument.
To avoid injury from electric shock or fire, and to
Warning:
prevent damage to the workstation and ancillary equipment,
do not place objects filled with liquid—such as solvent
bottles—on these items, or expose them to dripping or
splashing liquids.
Use shielded network cables with the instrument to reduce
Do not connect the instrument's power supply cord until

Connecting the workstation

March 14, 2014, 715004387 Rev. B
235