B External Connections

Connecting to the nitrogen gas supply

Required materials
Chemical-resistant, powder-free gloves
Sharp knife
Wrench
6-mm PTFE tubing (included in the Waters Rough Pump Connect Kit)
Nitrogen regulator
To connect the nitrogen gas supply:
1.
Use the sharp knife to cut a 3.8-cm to 5.0-cm (1.5-inch to 2-inch) length
of 6-mm PTFE tubing.
Requirement:
horizontal axis).
2.
Connect one end of the 6-mm tubing to one end of the nitrogen supply
in-line filter.
3.
Connect the remaining length of the 6-mm PTFE tubing to the other end
of the filter.
4.
Connect the free end of the short piece of 6-mm PTFE tubing to the
nitrogen inlet port on the rear of the instrument.
228
March 14, 2014, 715004387 Rev. B
Cut the tubing squarely (that is, perpendicularly to its