Fitting the extraction cone to the ion block
Required materials
•
Chemical-resistant, powder-free gloves
•
Combined 2.5-mm Allen wrench and cone extraction tool
Warning:
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the ion block.
To fit the extraction cone onto the ion block:
1.
Fit the extraction cone handle insulator onto the extraction cone handle.
2.
Fit the extraction cone onto the ion block.
3.
Rotate the 2 PEEK extraction cone retainer clips to secure the extraction
cone, then use the combined 2.5-mm Allen wrench and cone extraction
tool to tighten the retainer clip securing screws.
Fitting the ion block assembly to the source assembly
Required materials
•
Chemical-resistant, powder-free gloves
•
Combined 2.5-mm Allen wrench and cone extraction tool
Warning:
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the source components.
Warning:
working with the source enclosure open if one or both of these
conditions apply:
• An ESI probe is fitted (the probe tip is sharp).
• A corona pin is fitted (the pin tip is sharp).
To avoid personal contamination with biohazards or
To avoid personal contamination with biohazards or
To avoid puncture wounds, take great care while
March 14, 2014, 715004387 Rev. B
Cleaning the extraction cone
125