2 Preparing the Mass Spectrometer for Operation
Warning:
toxic compounds, always wear chemical-resistant, powder-free
gloves while performing this procedure.
To install the low-volume vials:
1.
If a standard reservoir bottle is fitted, remove it.
2.
Screw the low-volume adaptors into the manifold, and finger tighten
them.
3.
Screw the low-volume vials into the adaptors.
Purging the infusion pump
Whenever you replace a solution bottle, purge the infusion pump with the
solution that you are going to use next. See the mass spectrometer's online
Help for details.
Depending on the solutions used, the instrument's solvent delivery
Tip:
system can require more than one purge cycle to minimize carryover.
48
February 11, 2013, 71500130002 Rev. D
To avoid becoming contaminated with biohazards or
Low-volume adaptor
Low-volume vial