Warning:
or compounds that are toxic or corrosive, and to avoid
spreading contamination to uncontaminated surfaces, wear
clean, chemical-resistant, powder-free gloves when working
with the source components.
To remove the probe adjuster assembly probe and source enclosure seals:
1.
Remove the source enclosure from the instrument (see
source enclosure from the instrument" on page
2.
Use the O-ring removal kit to carefully remove the following seals from
the probe adjuster assembly:
Probe seal
Nebulizer gas seal
See
"Removing O-rings and seals" on page
Replacing the source assembly seals
To avoid personal contamination with biohazards
Probe adjuster nebulizer gas seal
Probe adjuster assembly probe seal
February 11, 2013, 71500130002 Rev. D
"Removing the
87).
97.