Rebooting the mass spectrometer
The reset button shuts down the electronics momentarily and causes the mass
spectrometer to reboot.
Reboot the mass spectrometer when either of these conditions applies:
•
The mass spectrometer software fails to initialize.
•
Immediately following a software upgrade.
To reboot the mass spectrometer:
1.
Open the mass spectrometer's front, left-hand door.
2.
Insert a short length (7.5 cm) of PEEK™ tubing, or similar object, into
the reset button aperture to operate the reset button.
3.
Remove the PEEK tubing from the reset button aperture.
4.
Close the mass spectrometer's door.
Rebooting the mass spectrometer
Reset button aperture
February 11, 2013, 71500130002 Rev. D
49