Assembling The Sampling Cone Assembly - Waters Xevo TQ MS Operator's, Overview And Maintenance Manual

Hide thumbs Also See for Xevo TQ MS:
Table of Contents

Advertisement

Assembling the sampling cone assembly

Required materials
Chemical-resistant, powder-free gloves
Caution:
• To avoid recontaminating the sampling cone assembly, wear clean
chemical-resistant, powder-free gloves during this procedure.
• To avoid damaging the fragile sample cone, do not place it on its tip;
instead, place it on its flanged base.
To assemble the sampling cone assembly:
1.
Fit the cone gas nozzle handle to the cone gas nozzle and turn the handle
clockwise to tighten.
O-ring
Sample cone
Cone gas nozzle
Cone gas nozzle handle
2.
Carefully fit the sample cone into the cone gas nozzle.
3.
Fit the O-ring into the groove created between the sample cone and cone
gas nozzle. (Fit a new O-ring if the old one has been disposed of.)
Cleaning the sampling cone assembly