Caution:
push the bulb out of the lamp drive assembly.
6.
Pull the bulb out of the lamp drive assembly.
7.
Insert the new bulb into the lamp drive assembly.
The lamp glass is magnesium fluoride. Avoid touching it because
Tip:
dirt or other contaminants on the window significantly reduce UV
transmission.
8.
Refit the lamp-bulb access plug.
9.
Close the source enclosure.
10. Slide closed the instrument's source interface door.
APPI/APCI source—changing the UV lamp bulb
To avoid breaking the bulb, do not use a screwdriver to
February 11, 2013, 71500130002 Rev. D
183