Required materials
Chemical-resistant, powder-free gloves
To empty the nitrogen exhaust trap bottle:
1.
To stop the LC flow, follow the action below for your software:
Software
MassLynx
UNIFI
2.
Pull the source enclosure release (located at the bottom, right-hand side)
outwards, and swing open the enclosure.
Warning:
or toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves while handling the nitrogen exhaust
trap bottle. The waste liquid in the nitrogen exhaust trap
bottle comprises ACQUITY UPLC solvents and analytes.
3.
Unscrew and remove the nitrogen exhaust trap bottle from the cap and
associated fittings.
Warning:
waste liquid according to local environmental regulations.
The waste liquid can be contaminated with biohazards or
toxic materials.
4.
Dispose of the waste liquid in accordance with local environmental
regulations.
5.
Fit and tighten the nitrogen exhaust trap bottle to the cap.
6.
Secure the nitrogen exhaust trap bottle in the upright position.
7.
Close the source enclosure.
An automatic leak test will now be performed.
Tip:
Action
In the instrument console, click Stop Flow
On the System Console tool bar, click Stop Flow
To avoid personal contamination with biohazards
To avoid spreading contamination, dispose of
February 11, 2013, 71500130002 Rev. D
Emptying the nitrogen exhaust trap bottle
.
.
99