4 Maintenance Procedures
3.
Fit the UNF coupling to the new capillary.
4.
Use the needle-nose pliers to slide a new liner sleeve and ferrule onto
the capillary.
5.
Insert the capillary in the PEEK union, and ensure that it is fully seated.
6.
Screw the UNF coupling into the PEEK union, finger-tight only.
7.
Gently tug on the capillary, testing to ensure that it stays in place.
8.
Use the 7-mm wrench to tighten the locknut against the PEEK union
until the union can no longer be twisted.
9.
Slide a new conductive sleeve and the knurled collar over the capillary.
10. Tighten the knurled collar to the UNF coupling.
Warning:
goggles when performing the leak test.
11. Perform a leak test by attaching the free end of the PEEK tubing to an
LC pump and pumping 50:50 acetonitrile/water through it, at 1 mL/min.
If leakage occurs, disassemble and remake the connection, and
repeat the leak test.
If the backpressure on the LC pump is high, replace the capillary,
and repeat the leak test.
12. When no leakage occurs and the backpressure on the LC pump is
normal, disconnect the PEEK tubing from the LC pump.
156
February 11, 2013, 71500130002 Rev. D
Probe inlet connector
PEEK tubing
To avoid high-pressure liquid jet spray, wear safety
TP02671