Waters Xevo TQ MS Operator's, Overview And Maintenance Manual page 71

Hide thumbs Also See for Xevo TQ MS:
Table of Contents

Advertisement

Caution:
breaking, always retract the stage before installing the source
enclosure or closing the door.
4.
On the NanoFlow source, release the stage retaining-screw, pull the stop
screw and slide the stage fully out of the enclosure.
Stop screw
5.
Using both hands, fit the NanoFlow source enclosure to the two
supporting studs on the source adaptor housing.
6.
Close the source enclosure door.
7.
Connect a 1/16 inch PTFE tube between the mass-flow controller output
(mounted beneath the stage on the front of the NanoFlow source) and
your sprayer.
For details regarding how to fit each sprayer, see the corresponding
Tip:
reference:
Waters Universal NanoFlow Sprayer Installation and Maintenance
Guide (part number 71500110107)
"Fitting a borosilicate glass capillary (nanovial)" on page 73
Capillary Electrophoresis/ Capillary Electrochromatography
Sprayer User's Guide (part number 6666522)
To prevent the sprayer colliding with the cone and
February 11, 2013, 71500130002 Rev. D
NanoFlow ESI source