Waters Xevo TQ MS Operator's, Overview And Maintenance Manual page 134

Hide thumbs Also See for Xevo TQ MS:
Table of Contents

Advertisement

4 Maintenance Procedures
18. Use the combined 2.5-mm Allen wrench and cone extraction tool to
remove the captive screws securing the 2 PEEK extraction cone retainer
clips.
Caution:
• To avoid damaging the extraction cone aperture, take great
• To avoid damaging the fragile extraction cone, do not place it
19. Remove the extraction cone from the ion block.
134
February 11, 2013, 71500130002 Rev. D
care when removing the extraction cone from the ion block.
on its tip; instead, place it on its flanged base.
Extraction cone aperture
Securing screw
Retainer clip
Extraction cone

Advertisement

Table of Contents
loading

Table of Contents