C External Connections

Connecting the workstation

Before connecting the workstation to the instrument, set up the workstation
according to its accompanying instructions. Locate the workstation within
5 meters (16 feet) of the instrument.
Requirement:
ensure compliance with FCC limits.
To connect the workstation:
1.
Connect the monitor to the PC.
2.
Connect one end of the shielded network cable to the appropriate port on
the rear panel of the mass spectrometer.
3.
Connect the other end of the shielded network cable to the port labeled
instrument LAN on the workstation rear panel.
To connect the instrument to the power source:
Requirement:
you complete the installation procedures in the previous sections.
1.
Select the appropriate power cord for your location.
2.
Connect the female end of the power cord to the power port on the rear
panel of the instrument.
236
February 11, 2013, 71500130002 Rev. D
To avoid injury from electric shock or fire, and to
Warning:
prevent damage to the workstation and ancillary equipment,
do not place objects filled with liquid—such as solvent
bottles—on these items, or expose them to dripping or
splashing liquids.
Use shielded network cables with the mass spectrometer to
Do not connect the instrument's power supply cord until