4 Maintenance Procedures

Fitting the sampling cone assembly to the source

Required material
Chemical-resistant, powder-free gloves
Warning:
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the source components.
Warning:
working with the source enclosure open if one or both of these
conditions apply:
• An ESI probe is fitted (the probe tip is sharp).
• A corona pin is fitted (the pin tip is sharp).
Caution:
do not open the source isolation valve before fitting the sampling
cone assembly to the ion block assembly.
118
February 11, 2013, 71500130002 Rev. D
To avoid personal contamination with biohazards or
To avoid puncture wounds, take great care while
To avoid damaging the instrument by sudden venting,