Waters Xevo TQ MS Operator's, Overview And Maintenance Manual page 74

Hide thumbs Also See for Xevo TQ MS:
Table of Contents

Advertisement

3 Changing the Mode of Operation
5.
Unscrew the union from the end of the sprayer assembly.
6.
Remove the existing capillary from the sprayer.
7.
Carefully remove the new borosilicate glass capillary from its case by
lifting vertically while pressing down on the foam with two fingers.
Foam
8.
Load sample into the capillary using either a fused silica syringe needle
or a GELoader
and the sample.
74
February 11, 2013, 71500130002 Rev. D
®
tip, minimizing any bubbles between the capillary tip
Capillary
Union
Capillary