Bear in mind these requirements when connecting the roughing pump:
The pump must be horizontal, or within one degree of horizontal.
The ambient temperature of area in which you place the pump must
range between 12 and 40 ºC (54 to 104 ºF).
To ensure proper ventilation, install the pump with the following
minimum clearances:
Back side minimum clearance
is 15.24 cm (6 inches)
Left side minimum clearance
is 15.24 cm (6 inches)
Connecting the Edwards oil-free roughing pump
Right side minimum clearance
is 15.24 cm (6 inches)
TP03165
Front side minimum clearance
is 35.56 cm (14 inches)
February 11, 2013, 71500130002 Rev. D
.
225