Bear in mind these requirements when connecting the roughing pump:
The pump must be horizontal, or within one degree of horizontal.
The ambient temperature of the area in which you place the pump must
range between 12 and 40 ºC (54 to 104 ºF).
The pump must be oriented so that it permits easy access to the oil
exhaust filter and oil-level sight glass.
For proper ventilation, the following minimum clearances must apply:
Left side minimum clearance
is 15.24 cm (6 inches)
Front side minimum clearance
is 35.56 cm (14 inches)
Warning:
or toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the pumps and their
connections.
Connecting the Varian oil-filled roughing pump
TP03159
Right side minimum clearance
is 15.24 cm (6 inches)
To avoid personal contamination with biohazards
February 11, 2013, 71500130002 Rev. D
Back side minimum clearance
is 15.24 cm (6 inches)
219