Purging The Infusion Pump - Waters Xevo TQ MS Operator And Maintenance Manual

Table of Contents

Advertisement

To install the low-volume vials
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
1.
If a standard reservoir bottle is fitted, remove it.
2.
Screw the low-volume adaptors into the manifold, and finger tighten
them.
3.
Screw the low-volume vials into the adaptors.

Purging the infusion pump

Whenever you replace a solution bottle, purge the infusion pump with the
solution that you are going to use next. See the mass spectrometer's online
Help for details.
Depending on the solutions used, the instrument's solvent delivery
Tip:
system can require more than one purge cycle to minimize carryover.
The reservoir bottles can be contaminated with
Preparing the IntelliStart Fluidics system
Low-volume adaptor
Low-volume vial
2-7