2.
Use the 3-mm Allen wrench to unscrew and remove the 4 screws
securing the PEEK ion block support to the adaptor housing.
Adaptor housing
PEEK ion block
support
Securing screws
3.
Remove the PEEK ion block support from the adaptor housing.
4.
Use the O-ring removal kit to carefully remove all the O-rings from the
PEEK ion block support (see
5.
If any of the O-rings show signs of deterioration or damage, dispose of
them in accordance with local environmental regulations.
Caution:
from the adaptor housing:
• avoid scratching the internal surfaces of the adaptor block.
• do not compress the hexapole rods.
6.
Carefully grasp the source hexapole assembly and remove it from the
adaptor housing.
"Removing O-rings and seals" on page
To avoid contaminating uncontaminated
Warning:
surfaces with biohazards or compounds that are toxic
or corrosive, dispose of the O-rings according to local
environmental regulations.
To avoid damage when removing the hexapole assembly
February 11, 2013, 71500130002 Rev. D
Cleaning the source hexapole assembly
Hexapole
assembly
97).