Waters Xevo TQ MS Operator's, Overview And Maintenance Manual page 153

Hide thumbs Also See for Xevo TQ MS:
Table of Contents

Advertisement

6.
Remove the metal gasket from the probe tip.
Metal gasket
7.
Remove the PEEK union/UNF coupling assembly and capillary from the
probe.
PEEK union/UNF coupling assembly
8.
Unscrew and remove the knurled collar from the UNF coupling.
PEEK union
Locknut
Conductive sleeve
9.
Remove the knurled collar and conductive sleeve from the capillary.
Capillary
UNF coupling
Knurled collar
February 11, 2013, 71500130002 Rev. D
Replacing the ESI probe sample capillary
153