4 Maintenance Procedures
5.
Fit the mounting shaft insertion tool to the mounting shaft.
Caution:
the mounting shaft insertion tool to the mounting shaft before
fitting the shaft to the lamp mounting flange.
6.
Slide the lamp mounting flange back over the shaft taking care to get
the correct orientation.
7.
Reinsert that shaft through the side of the source enclosure, and fit the
lamp mounting flange to the APPI source enclosure side.
Requirement:
increments until they are all fully tight. Doing so ensures that the lamp
mounting flange is uniformly seated on the APPI source enclosure side
plate.
8.
Use the 2.5-mm Allen wrench to tighten the four mounting-flange
securing screws.
9.
Remove the mounting shaft insertion tool from the mounting shaft.
10. Fit the PEEK insulator to the end of the mounting shaft.
11. Use the 2.5-mm Allen wrench to fit and tighten the two insulator
retaining screws.
12. Fit the repeller electrode to the PEEK insulator.
192
February 11, 2013, 71500130002 Rev. D
To prevent damage to the mounting shaft O-rings, fit
Tighten the securing screws sequentially and by small
Mounting shaft insertion tool
XMS0017