5.
Remove the 2 source enclosure, release-handle screws, and remove the
handle.
4 source-enclosure
cover screws
lamp-drive cover
screws (2)
Release handle
6.
Remove the remaining 2 screws, which are hidden by the release handle.
7.
Slide the cover off the lamp-drive.
8.
Remove the 4 source-enclosure cover screws.
9.
Ease the source-enclosure cover over the lamp drive assembly.
10. Use the 2.5-mm Allen wrench to unscrew the 4 mounting-flange screws.
APPI/APCI source—Replacing the APPI lamp drive seals
February 11, 2013, 71500130002 Rev. D
TP03204