4 Maintenance Procedures
Caution:
• To avoid damaging the probe heater's electrical wiring, take
• To avoid damaging the probe heater's electrical connections,
4.
Gripping the probe heater as shown, carefully pull it off the probe
assembly.
5.
Dispose of the probe heater in accordance with local environmental
regulations.
168
February 11, 2013, 71500130002 Rev. D
great care when gripping and handling the heater.
do not twist the heater when removing it from the probe
assembly.
Probe heater
To avoid spreading contamination, dispose of the
Warning:
probe heater according to local environmental regulations.
It can be contaminated with biohazardous, or toxic
materials.