Waters Xevo TQ MS Operator's, Overview And Maintenance Manual page 143

Hide thumbs Also See for Xevo TQ MS:
Table of Contents

Advertisement

4.
Add 1:1 methanol/water to the glass vessel until the hexapole assembly
is immersed completely.
5.
Place the vessel in the ultrasonic bath for 30 minutes.
Caution:
clean, chemical-resistant, powder-free gloves for the rest of this
procedure.
6.
Carefully remove the hexapole assembly from the vessel, and blow-dry it
using inert, oil-free gas.
7.
Inspect the hexapole assembly for persisting contamination, If
contamination is present, do as follows:
a.
Use the wash bottle containing methanol to rinse the source
hexapole assembly over the large beaker.
b.
Blow-dry the hexapole assembly with inert, oil-free gas.
8.
Use the small flat-blade screwdriver to ensure that the hexapole
assembly screws are tight.
Hook
Rear support ring
TP02658
To avoid recontaminating the hexapole assembly, wear
February 11, 2013, 71500130002 Rev. D
Cleaning the source hexapole assembly
143