13. Use the small Phillips screwdriver to fit and tighten the repeller
electrode retaining screw.
14. Insert the UV bulb into the lamp drive assembly and push it fully home.
15. Fully retract the lamp mounting shaft from the source enclosure.
16. Refit the lamp-assembly collar-cover, and secure it on its base with the 4
screws.
17. Refit the lamp assembly cover, and secure it on its base (2 screws) and
above the bulb extraction aperture (2 screws).
18. Refit the source enclosure release handle, and secure it with the 2
screws.
19. Refit and reconnect the source-enclosure to the machine, refit the corona
pin and probe. Refer to
page 66
for instructions.
APPI/APCI source—Replacing the APPI lamp drive seals
"To install the combined APPI/APCI source:" on
February 11, 2013, 71500130002 Rev. D