Fitting The Extraction Cone To The Ion Block - Waters Xevo G2 QTof Operator's, Overview And Maintenance Manual

Hide thumbs Also See for Xevo G2 QTof:
Table of Contents

Advertisement

5.
Inspect the extraction cone for persisting contamination. If
contamination is present, do as follows:
a.
Use the wash bottle containing 1:1 methanol/water to rinse the
extraction cone over the large beaker.
b.
Blow dry the extraction cone with inert, oil-free gas.
6.
Inspect the extraction cone for persisting contamination.
Requirement:
procedure. If contamination is still present, dispose of the extraction
cone, and obtain a new one.

Fitting the extraction cone to the ion block

Required materials
Chemical-resistant, powder-free gloves
Combined 2.5-mm Allen wrench and cone extraction tool
To fit the extraction cone to the ion block:
1.
Fit the extraction cone handle insulator to the extraction cone handle.
2.
Fit the extraction cone to the ion block.
3.
Rotate the 2 PEEK extraction cone retainer clips to secure the extraction
cone, and then use the combined 2.5-mm Allen wrench and cone
extraction tool to tighten the retainer clip securing screws.
5-44
Maintenance Procedures
The extraction cone can be contaminated with
Warning:
biohazardous and/or toxic materials. Dispose of it according
to local environmental regulations.
If contamination is present, repeat the cleaning
The ion block can be contaminated with biohazardous
Warning:
and/or toxic materials. Always wear chemical-resistant,
powder-free gloves while performing this procedure.

Hide quick links:

Advertisement

Table of Contents
loading