Installing The Apgc Ion Chamber On The Mass Spectrometer's Ion Block - Waters APGC Operator's Manual Supplement

Atmospheric pressure gas chromatography
Table of Contents

Advertisement

Installing the APGC ion chamber on the mass spectrometer's ion
block
Required materials
Chemical-resistant, powder-free gloves
Combined, 2.5-mm Allen wrench and cone extraction tool
Dentist's mirror
Requirement:
mass spectrometer, you must use the reduced-aperture, 0.36-mm, sample cone
supplied with the APGC, to avoid problems associated with unstable signal
intensity.
To install the APGC ion chamber:
1.
Select the correct sample cone for the instrument you intend to use with
the APGC:
For Xevo TQD or SQ Detector 2 instruments, use the 0.36-mm
sample cone provided with the APGC. The cone is labelled "0.36mm"
on its base and side wall.
For all other instruments, use the sample cone that you retained in
the previous procedure
2.
Ensure that the O-ring is in position on the sample cone.
To avoid personal contamination with biologically
Warning:
hazardous, toxic, or corrosive materials, and to avoid
spreading contamination to uncontaminated surfaces, wear
clean, chemical-resistant, powder-free gloves when installing
the ion chamber.
To avoid burn injuries, take great care while
Warning:
working with the source components, which can be hot.
During APGC operation with the Xevo TQD or SQ Detector 2
Preparing the mass spectrometer for operation with the APGC
(page
2-8,
step
8)
2-9