From the table below, select the use of the alarm output function.
Display
0 pps
Measured value
Last valid
Max pps
5.14.3 Pulse Width Setting
The pulse width can be selected to use the pulse output.
The setting can be configured with the following parameter.
Menu path
Display
F
OUNDATION
Fieldbus
Block
Relative
Name
Index
MTB
104
From the table below, select the pulse width.
Display
0.05 ms
0.1 ms
0.5 ms
1 ms
20 ms
33 ms
50 ms
100 ms
200 ms
330 ms
500 ms
1000 ms
2000 ms
*
Duty cycle 50%
*:
The maximum pulse width is up to 300 seconds (5 minutes) when "Duty cycle 50%" is selected.
When outputting the pulse whose period is over 10 minutes, the pulse width is fixed at 300s.
Selection
F
OUNDATION
Fieldbus
0 pps
Measured value
Last valid
Max pps
Device setup ► Detailed setup ► Pulse/Status out ► PO1/SO1 ► (see below)
Device Configuration ► MTB ► Device Configuration ► Maintenance ► Pulse/Status Out
► (see below)
Display
Fix width
Selection
F
0.05ms
0.1ms
0.5ms
1ms
20ms
33ms
50ms
100ms
200ms
330ms
500ms
1000ms
2000ms
Duty cycle 50%
Pulse output stop
Outputs the frequency or fixed pulse with the measured value
calculated while an alarm occurs.
Outputs the frequency or pulse with the last instantaneous value
right before an alarm occurs (output is monotonic increase).
Outputs the frequency or pulse at 12500 pps.
Parameter
F
Fieldbus
OUNDATION
Pulse 1 Fix Width
Fieldbus
OUNDATION
*
<5. Functions>
Description
Description
Specifies the pulse width.
Max. output frequency [Hz]
10000
5000
1000
500
25
15
10
5
2.5
1.5
1.0
0.5
0.25
-
IM 01E21A02-03EN
165