Autozero function
executing...
SET
SFT
INC
F0407.ai
Autozero
Execute
▲
Result
▼
SET
SFT
INC
F0408.ai
When executing zero adjustment from F
procedures of the interactive operation guide called DD Method. Follow procedures on the
operation screen.
Confirmation of zero adjustment result
The zero adjustment result using the display can be confirmed with the following parameter.
Menu path
Display
F
OUNDATION
Fieldbus
Block
Relative
Name
Index
STB
60
Result
Zero value
▲
▼
SET
SFT
INC
F0409.ai
Zero value
– 00.018cm/s
–
00.018cm/s
– 99.999
+ 99.999
SET
SFT
INC
F0410.ai
NOTE
When the zero adjustment result exceeds the defined value, the warning [092:AZ warn] is
displayed.
The zero adjustment starts, and the progress is displayed with the remaining time and a
00 : 00
bar graph. Wait for the zero adjustment to complete.
The time remaining
00 : 08
until the end.
After the zero adjustment is finished, the display returns to "Autozero" menu.
00 : 00
►
Device setup ► Diag/Service ► Autozero ► Result ► (see below)
Device Configuration ► STB ► Device Configuration ► Maintenance ► Autozero ► (see
below)
Display
Zero value
For the result of the zero adjustment, select "Result" and then "Zero value".
00 : 00
The result of zero adjustment is displayed as on the left of the screen.
00 : 00
Fieldbus, zero adjustment is set with the
OUNDATION
Parameter
F
Fieldbus
OUNDATION
Zero Value
<5. Functions>
Description
Refer to the following
confirmation methods for
results of zero adjustment and
the display.
IM 01E21A02-03EN
65