5.2.6

Totalization Direction Setting

The totalization direction can be specified when using the totalization function.
The setting can be configured with the following parameters.
Menu path
Display
F
OUNDATION
Fieldbus
Block
Relative
Name
Index
STB
70
STB
79
From the table below, select the totalization direction.
Selection
Display
Balanced
Absolute
Only positive
Only negative
Hold
5.2.7

Totalized Value Reset/Preset Function

The reset/preset function can be specified for the totalized value. If the reset function is used,
the function resets the totalized value to "0". If the preset function is used, it sets the preset value
specified in advance to the totalized value. The preset function is used when starting to count
totalization from the specified value.
The setting can be configured with the following parameters.
Use of the reset/preset function
Menu path
Display
F
OUNDATION
Fieldbus
Block
Relative
Name
Index
STB
72
STB
81
Device setup ► Detailed setup ► Totalizer ► (see below)
Device Configuration ► STB ► Device Configuration ► Configuration ► Totalizer ► (see
below)
Parameter
Display
Totalizer 1 ► Options Totalizer1 ► Total 1 Options
Totalizer 2 ► Options Totalizer2 ► Total 2 Options
F
OUNDATION
Fieldbus
Balanced
Totalizes the flow rate (difference) of the forward and reverse directions.
Absolute
Totalizes the absolute value of the flow rate.
OnlyPositive
Totalizes only the flow rate in the forward direction.
OnlyNegative
Totalizes only the flow rate in the reverse direction.
Hold
Stops totalization processing (holding the current totalized value).
Device setup ► Detailed setup ► Totalizer ► (see below)
Device Configuration ► STB ► Device Configuration ► Configuration ► Totalizer ► (see
below)
Display
Totalizer 1 ►
Reset/Preset
Totalizer 2 ►
Reset/Preset
F
Fieldbus
OUNDATION
Description
Parameter
F
Fieldbus
OUNDATION
Totalizer1 ► Total 1 Preset
Totalizer2 ► Total 2 Preset
<5. Functions>
Description
Specifies the totalization
direction of totalizer 1.
Specifies the totalization
direction of totalizer 2.
Description
Uses the reset/preset
function of totalizer 1.
Uses the reset/preset
function of totalizer 2.
IM 01E21A02-03EN
69