Device Information
Sensor Serial No
Main Board Software Rev.
Sensor Board Software Rev.
Display Board Software Rev.
Model (Note)
Communication and I/O Code
NOTE
Data can only be duplicated or restored between products of the same model code.
The setting can be configured with the following parameters.
For details about the restorable parameters, refer to Subsection 5.11.3.
Menu path
Display
Device setup ► Diag/Service ► Param bkup/restore ► (see below)
F
Device Configuration ► MTB ► Device Configuration ► Maintenance ► Parameter
OUNDATION
Fieldbus
Backup/Restore ► (see below)
Block
Relative
Name
Index
MTB
83
Restore execute
MTB
84
Restore result
*1:
From the table below, select the execution of the restore function. In the communication access of F
Fieldbus, the communication access is set with the procedures of the interactive operation guide called DD Method.
Selection
Display
F
OUNDATION
Not execute
Not execute
Duplicate data 1
Duplicate Data1
Duplicate data 2
Duplicate Data2
Duplicate data 3
Duplicate Data3
Duplicate SD
Duplicate SD
Restore data 1
Restore Data1
Restore data 2
Restore Data2
Restore data 3
Restore Data3
Restore SD
Restore SD
Compulsion data 1
Compulsion Data1
Compulsion data 2
Compulsion Data2
Compulsion data 3
Compulsion Data3
Compulsion SD
Compulsion SD
Restore factory
Restore Factory
Duplicate Data
-
Parameter
Display
F
OUNDATION
Restore Execute
Restore Result
Fieldbus
Does not restore data.
Restores the parameter (Duplicate Data) setting from memory 1
on the display board to the main board of the product.
Restores the parameter (Duplicate Data) setting from memory 2
on the display board to the main board of the product.
Restores the parameter (Duplicate Data) setting from memory 3
on the display board to the main board of the product.
Restores the parameter (Duplicate Data) setting from the
microSD card to the main board of the product.
Restores the parameter (Restore Data) setting from memory 1
on the display board to the main board of the product.
Restores the parameter (Restore Data) setting from memory 2
on the display board to the main board of the product.
Restores the parameter (Restore Data) setting from memory 3
on the display board to the main board of the product.
Restores the parameter (Restore Data) setting from the microSD
card to the main board of the product.
Restores the parameter (Compulsion Data) setting from memory
1 on the display board to the main board of the product.
Restores the parameter (Compulsion Data) setting from memory
2 on the display board to the main board of the product.
Restores the parameter (Compulsion Data) setting from memory
3 on the display board to the main board of the product.
Restores the parameter (Compulsion Data) setting from the
microSD card to the main board of the product.
Restores to the status that is set at shipment from the
manufacturing factory.
<5. Functions>
Restore Data
Description
Fieldbus
Specifies the execution of the
restore function.
Displays the result of the
restore function.
Description
150
Compulsion Data
-
-
-
-
-
-
*1
*2
OUNDATION
IM 01E21A02-03EN