YOKOGAWA Admag TI Series User Manual page 152

Foundation fieldbus communication type
Hide thumbs Also See for Admag TI Series:
Table of Contents

Advertisement

*2:
The result of the restore function is displayed as shown below.
Selection
Display
Unexecuted
Success
Failure
Running
The table below shows alarms whose parameters can be restored or duplicated when an alarm
occurs.
Display
010:Main CPU FAIL
011:Rev calc FAIL
012:Main EEP FAIL
013:Main EEP dflt
014:Snsr bd FAIL
015:Snsr comm ERR
016:AD 1 FAIL[Sig]
017:AD 2 FAIL[Excit]
018:Coil open
019:Coil short
020:Exciter FAIL
027:Restore FAIL
028:Ind bd FAIL
029:Ind bd EEP FAIL
030:LCD drv FAIL
031:Ind bd mismatch
032:Ind comm ERR
033:microSD FAIL
050:Signal overflow
051:Empty detect
053:Adh over lv 4
060:Span cfg ERR
066:Density cfg ERR
067:Pls 1 cfg ERR
069:Nomi size cfg
070:Adh cfg ERR
071:FLN cfg ERR
072:Log not start
082:Pls 1 saturate
085:Cable miscon
086:Coil insulation
087:Adhesion lv 3
088:LC warn
089:Insu detect
090:FLN over lv 3
091:FLN over lv 4
092:AZ warn
093:Verif warn
094:Fact noise warn
095:Simulate active
098:Pls 1 fix
101:Param restore run
F
OUNDATION
Fieldbus
Unexecuted
Does not restore data.
Success
Succeeded in restoration.
Failure
Failed in restoration.
Running
Restore running
Alarm Name
10:Main board CPU failure
11:Reverse calculation failure
12:Main board EEPROM failure
13:Main board EEPROM default
14:Sensor board failure
15:Sensor communication error
16:A/D1 failure[Signal]
17:A/D2 failure[Exciter]
18:Coil open
19:Coil short
20:Exciter failure
27:Parameter restore incomplete
28:Indicator board failure
29:Indicator board EEP failure
30:LCD driver failure
31:Indicator board mismatch
32:Indicator communication error
33:microSD failure
50:Signal overflow
51:Empty pipe detection
53:Adhension over level 4
60:Span configuration error
66:Density configuration error
67:Pulse output 1 cfg error
69:Nominal size cfg error
70:Adhesion configuration error
71:Flow noise cfg error
72:Data logging not started
82:Pulse output 1 saturated
85:Cable misconnect
86:Coil insulation warning
87:Adhesion over level 3
88:Low conductivity Warning
89:Insulation detection
90:Flow noise over level 3
91:Flow noise over level 4
92:Autozero warning
93:Verification warning
94:Factory noise warning
95:Simulation active
98:Pulse output 1 fixed
101:Parameter restore running
<5. Functions>
Description
: Executable when an alarm occurs.
-: Not executable when alarm occurs.
F
Fieldbus
OUNDATION
151
Restore or
Duplicate
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
IM 01E21A02-03EN