Setup 1/2

This setup obtains the I-V Converter scaling factors.
Figure 2-2. Channel 1 Adjustment, Setup 1/2 (using DC Power Supply)
Perform the following steps as the program instructs.
1. Connect the 16340A furnished Triaxial Cable to the 4349B \Ch 1" connector.
2. Connect the other end of the Triaxial Cable to the Triaxial Tee Adapter (f).
3. Connect the 16340A furnished Triaxial-BNC(f) Adapter to the Triaxial Tee Adapter (f).
4. Connect the Triaxial Tee Adapter (m) directly to the 16340A's \10 10 - 10 2 " triaxial connector.
5. Connect the 16340A's \10 10 - 10 2 " BNC connector to the 16340A furnished BNC Tee Adapter
(f) using the 16340A furnished BNC cable.
6. Connect the BNC Tee Adapter (m) to the DC Power Supply voltage output using appropriate
adapter and cable.
7. Connect the BNC Tee Adapter (f) to the Multimeter input using the Dual Banana-BNC Cable.