Inoblender Test - Mallinckrodt INOMAX DSir Plus Operation Manual

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

INOblender Test

2
3
40 ppm
5
1
Acceptable
4
NO Value
• If the INOmax DS
WARNING:
depressurize the regulator supply line (see next page "Depressurizing the
Regulator Supply Line").
• If the INOmax DS
repeat purge procedure (see page 2-5).
• If the INOmax DS
use procedure (see page 2-1).
The INOmax DS
IR
Proceed to Section 3/ Patient Application.
Part No. 20717 Rev-01
2014-07
4
3
10 L/min
5
2
32-48 ppm
is not going to be used on a patient within 10 minutes,
IR
is not used and is pressurized for more than 10 minutes,
IR
is depressurized and not used within 12 hours, repeat pre-
IR
is now ready to connect to the patient.
1. Turn O
flowmeter OFF and remove pre-
2
use oxygen tubing from O
connect to front of INOblender.
2. Remove the injector module from the
pre-use set-up and reconnect the
adapters.
3. On the INOblender, set the INOMAX dose
to 40 ppm and INOblender flowmeter to
10 L/min.
4. Allow monitored NO value to stabilize and
then verify acceptable NO value on the
INOmax DS
.
IR
5. Turn the dose and oxygen flow to zero.
Remove the pre-use set-up from the
INOblender.
Pre-use checkout complete.
flowmeter and
2
2-9