Integrated Pneumatic Backup INOMAX Delivery Test

Performance Test

Set INOMAX Dose
Acceptable NO
Value
Acceptable NO
Value
2
FiO
2
Part No. 20717 Rev-01
2014-07
40 ppm
35-45 ppm
< 1.5 ppm
95% ± 3 %
1. Verify that the O
flowmeter is set to 10 L/min.
2
2. Turn the integrated backup INOMAX delivery to
ON (250 mL/min.).
Verify "Backup ON" alarm occurs.
3. Allow monitored values to stabilize (may take
up to 3 minutes) and make sure the NO and
NO
readings are within the following ranges:
2
NO = 14-26 ppm
4. Turn the integrated backup INOMAX delivery
to OFF.
1. Verify that the O
flowmeter is set to 10 L/min.
2
2. Set the INOMAX dose to 40 ppm, allow
values to stabilize.
3. Compare the INOmax DS
the values in the table.
4. Turn INOMAX dose to zero.
Allow 2-3 minutes for monitored values
Note:
to stabilize. If a monitored value is
outside the range indicated, do a high
range calibration for that sensor.
NO
≤ 1.0 ppm
2
monitor values to
IR
9-7

Advertisement