Ventilator Compatibility - Mallinckrodt INOMAX DSir Plus Operation Manual

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

8/ Product Specifications
• In the United States, the approved patient population for the INOmax DS
WARNING:
specified in the drug labeling for INOMAX
to term and near-term neonates with hypoxic respiratory failure. The INOmax
DS
• Outside of the United States, use of the INOmax DS
in accordance with INOMAX or INOflo, nitric oxide for inhalation prescribing
information as established with the national health authority.
• Patient disconnect and high-pressure alarms are required for the ventilator.
The INOmax DS
is compatible with most types of ventilators by connecting into the inspired limb of a patient's breathing circuit.
IR
The system measures the gas flow in the breathing circuit and then injects NO/N

Ventilator Compatibility

Ventilators/Breathing Systems validated for use in the United States
Manufacturer
Model
Baby-Plus Bubble
A-plus Medical
CPAP
Airon
pNeuton
Corporation
Bear
750ps (Cub)
Bio-Med
Crossvent 2
Devices
Bio-Med
Crossvent 4
Devices
Bio-Med
MVP-10
Devices
Bird
VIP
Part No. 20717 Rev-01
2014-07
is not intended to be used in other patient populations.
IR
Pediatric/
Neonatal
(nitric oxide) for inhalation, is limited
®
gas to produce the set NO concentration in ppm.
2
Inspiratory Flow Rate:
Respiratory Rate:
Airway Peak Pressure:
PEEP:
High
Transport
Adult
Frequency
, as
IR
is limited to the use
IR
Measure
Specification
L/min
2 - 120
bpm
6 - 60
cmH
O
0 - 70
2
cmH
O
0 - 20
2
Nasal
Continuous
High Flow
Positive
Anesthesia
Airway
Pressure
(CPAP)
Nasal
Cannula
8-1