Mallinckrodt INOMAX DSir Plus Operation Manual page 64

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

3-12
5. Open the cylinder valve on the new cylinder
(this may activate the "Two Cylinders Open"
alarm until the empty cylinder valve is closed).
When using the transport regulator/
Caution:
cap assembly (PN 10022) ensure
the cap is fully seated and in place
on the INOmeter and the infrared
cable is connected and latched
to the infrared connector port on
the back of the INOmax DS
Section 4/ Transport).
If using the INOmax DS
Note:
regulator/cap assembly, transfer the
cap from the exhausted INOMAX
cylinder to the new INOMAX cylinder
at this time (see page 4-2). "Cylinder
Not Detected" alarm may occur.
6. Close the cylinder valve on the empty
cylinder and remove the supply line from
the back of the INOmax DS
7. Depressurize by using the purge port on the
back of the INOmax DS
regulator from the empty cylinder.
Depressurize the regulator and
Caution:
hose by using the purge port on the
back of the INOmax DS
removing the regulator from the
INOMAX cylinder.
8. Load a new INOMAX cylinder onto the cart.
(see
IR
transport
IR
.
IR
and remove the
IR
before
IR
Part No. 20717 Rev-01
2014-07