Mallinckrodt INOMAX DSir Plus Operation Manual page 146

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

Indicator
26. Battery Failure
27. Low Calibration
28. Monitor Alarm Delay
Active
29. Patient Info Incomplete
30. Running on Battery
31. Set Dose is Zero,
Please Close Cylinder
Valve
5-16
Indicators
Possible Cause
a. Device cannot
communicate with battery.
a. Low calibration in progress.
A Low Calibration indicator
will display in the alarm
area of the screen during
the low calibration.
The Monitor Alarm Delay
Active indicator will be
displayed for two minutes:
a. Upon exit from the
calibration screen (whether
or not a calibration was
actually performed)
b. Following a calibration
c. Following completion of an
auto purge
Monitor alarms for O
, NO
2
and NO will be inactive
anytime the Monitor Alarm
Delay Active indicator is
displayed.
This delay only affects the
monitor alarms, all other
alarms remain active.
a. Patient identifier has not
been entered.
a. Device is operating on the
battery.
a. The set dose has been set
to zero and the INOMAX
cylinder valve is still open.
Recommended Action
Contact Technical Support.
No user action required.
No user action required.
No user action required.
No user action required.
,
2
Enter patient identifier.
1. Connect to AC main power source when
available (check main power indicator).
2. Make sure the power cord is fully inserted
into the Power Cord Inlet and that the power
cord clamp is secure.
Close the INOMAX cylinder valve and
depressurize regulator.
Part No. 20717 Rev-01
2014-07