4/ Transport
• It is recommended that a second (backup) transport regulator cap assembly is available
Caution:
during all transports.
• It is recommended that a second (backup) cylinder of INOMAX is available during all
transports.

Transport Options

A. When moving the INOmax DS
B. When removing INOmax DS
C. When using the INOblender as a stand-alone device (see Section C below).
D. When using a separate INOmax DS
A. Intrahospital transport (within the hospital) when moving the
INOmax DS
as a unit (cart and cylinders)
IR
1. Refer to cylinder duration chart as a guide to determine if there is enough drug to last through the
transport, including unexpected delays. Bring additional cylinders as appropriate.
2. Connect the INOblender oxygen hose to a 50 psig portable oxygen source.
3
Manually ventilate the patient using the INOblender while configuring the INOmax DS
transport ventilator (see figure 4-11 or 4-12).
4. Upon return:
a. Manually ventilate the patient while reattaching the INOmax DS
b. Confirm the operation of the INOmax DS
c. Reconnect the INOblender oxygen hose to a 50 psig wall oxygen source.
Part No. 20717 Rev-01
2014-07
as a unit (cart and cylinders), (see Section A below).
IR
and INOblender from the cart (see Section B below).
IR
and INOblender for transport (see Section D below).
IR
.
IR
to the
IR
to the bedside ventilator.
IR
4-1