Spontaneous Breathing Patient On A Nasal Cannula - Mallinckrodt INOMAX DSir Plus Operation Manual

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

Spontaneous Breathing Patient on a Nasal Cannula

The INOmax DS
can be used with nasal cannula to deliver INOMAX concentrations from 5-80 ppm and an
IR
oxygen flow rate as low as two L/min.
Conditioning of the oxygen flow prior to delivery through the injector module will help ensure the most
accurate flow measurement. Conditioning can be achieved by adding 300 mm of 22 mm hose between the
oxygen tubing and the Injector Module.
WARNING:
Do not use the integrated pneumatic backup with flow rates less than five
L/min.
1
2
3
4
1. O
Flowmeter
2
2. O
Tubing
2
3. 15M x 4.5 mm Adapter
4. 22M/15F x 22M/15F Adapter
5. 300 mm of 22 mm Hose
6. Injector Module
Figure 3-38 Example: Spontaneous Breathing Nasal Cannula Patient Circuit Diagram
3-46
10
11
12
5
4
6
7. O
8. Patient Nasal Cannula
9. Patient Gas Sample Line with Nafion
10. INOmax DS
11. Injector Module Electrical Cable
12. NO/N
5
4
3
7
Tubing Sample Tee
2
IR
Injector Tube
2
9
8
Part No. 20717 Rev-01
2014-07